The 5.5-axis closed-loop cryogenic sample holder provided by Fermion Instruments can achieve precise control of sample position and posture. It is mainly composed of four parts: cryogenic sample rod, temperature control system, XYZ-R1 translation stage, and motor control system. The entire sample holder and translation stage are made of non-magnetic materials and can be used under ultra-high vacuum. The sample stage can move over a large range in the X, Y, and Z directions, and can achieve rotational movement around the Z axis (R1). The sample holder head has added sample stage pitch angle rotation (R2) and in-plane angle rotation (R3) functions. The motor control software can be linked with ScientaOmicron's analyzer to achieve automapping function.
● Low operating cost: Adopt air (water) cooled GM refrigerator to provide cooling capacity; | ● Multiple degrees of freedom: X, Y, Z, R1, R2 can all be controlled by motors, and R3 is operated by a vacuum manipulator; | ● Convenient operation: Plug and play, can run stably for a long time; |
● Ultra-low temperature: The minimum stability of the sample stage can reach 5.5K; | ● Accurate temperature control: The PID temperature control accuracy can reach 10mK; | ● Rapid temperature change: From room temperature to 5.5K, less than 4 hours; |
R1 axial rotation | 360° (-180°, + 180°) |
R2 pitch rotation | 60° (-15°, + 45°) |
R3 in-plane angular rotation | 240° (-120°, +120°) |
X / Y itinerary | 25mm (±12.5mm) |
Z itinerary | Maximum 650mm |
4-Axis | 5A-Axis | 5F-Axis | 5.5-Axis | 6-Axis | ||
XY itinerary | ±12.5mm | ±12.5mm | ±12.5mm | ±12.5mm | ±12.5mm | |
Z travel | Up to 700mm | Up to 700mm | Up to 700mm | Up to 700mm | Up to 700mm | Up to 700mm |
XYZ drive | Manual/motor | Manual/motor | Manual/motor | Manual/Motor | ||
XYZ Accuracy | <0.01mm | <0.01mm | <0.01mm | <0.01mm | <0.01mm | |
R1 Range | 360° (-180°, +180°) | 360°(-180°,+180°) | 360°(-180°,+180°) | 360°(-180°,+180°) | 360°(-180°,+180°) | |
R1 control mode | Manual/motor | Manual/motor | Manual/motor | Manual/motor | Manual/motor | |
R1 control accuracy | <0.1° | <0.1° | <0.1° | <0.1° | <0.1° | |
R1 reduction ratio | 360:0.2 | 360:0.2 | 360:0.2 | 360:0.2 | 360:0.2 | |
R2 (Flip) range | —— | —— | 60° (-15°, +45°) | 60° (-15°, +45°) | 60° (-15°, +45°) | |
R2 driver | —— | —— | Motor | Manual/Motor | Manual/Motor | |
R2 Accuracy | —— | —— | <0.1° | <1° | <0.1° | |
R2 Reduction Ratio | —— | —— | 360:9 | 360:6 | 360:3 | |
R3 (Azi) Scope | —— | 240° (-120°, +120°) | —— | 240° (-120°, +120°) | 240° (-120°, +120°) | |
R3 drive | —— | Manual/motor | —— | Manual | Manual/motor | |
R3 accuracy | —— | <0.1° | —— | Manual<1° | <0.1° | |
R3 reduction ratio | —— | 360:5 | —— | 360:4.5 | 360:4 | |
Temperature range | <4-325K | <5-325K | <5-325K | <6-325K | <6-325K | |
Minimum temperature | <4K | <5K | <5K | <6K | <6K | |
Control accuracy | <0.01K | <0.01K | <0.01K | <0.01K | <0.01K | |
Cooling time | <3 Hour (RT-6K) | <4 Hour (RT-6K) | <4 Hour (RT-6K) | <5 Hour (RT-6K) | <6 Hour (RT-6K) |